100003646中文資料圖爾克數(shù)據(jù)手冊(cè)PDF規(guī)格書(shū)
100003646規(guī)格書(shū)詳情
The 1-channel temperature measuring amplifier IM12-TI02-1TCURTDR-1I-1RI-C0/24VDC
has inputs for: Thermocouples acc. to IEC
60584, DIN 43710, GOST R 8.585-2001, low
voltages (-150 … +150 mV), RTDs acc. to
IEC 60751, DIN 43760, GOST 6651-94 (2, 3
and 4-wire) and resistors 0…5 kò (2, 3 and 4-
wire).
The device is parameterized via FDT and
IODD with a PC. The cold junction compensation can be set to internal, external, or to a
constant value. The current output can be set
to 0/4…20 mA (source or sink optional). In accordance with the parameterization, the input
signals are provided as a 0/4…20 mA standard current signal. With the changeover relay, either a limit value can be monitored for
excess or undershooting or a range can be
monitored.
The devices have a green power LED (Pwr)
and a red LED to indicate internal faults. For
each input circuit there is a yellow and a red
status LED. A fault in the input circuit leads
to a flashing red LED according to NE44, an
internal error to a steady red LED. The fault
current can be adjusted to 21.5
mA.
The device can be used in safety circuits up
to SIL2 (high and low demand according to
IEC 61508) and meets the requirements of
the NE21. It is equipped with removable screw
terminals.
The device is equipped with removable screw
terminals.
■ Input circuits monitored for wire-break and
short-circuit
■ Parameterized via PC
■ Complete galvanic isolation
■ Removable screw terminals
■ ATEX use in Zone 2, cUL
■ SIL 2